OCM-510 Series

A spectrometer designed for precise measurement of light source quality and quantity.

Overview

OCM-510 Spectroradiometer is equipped with a high-performance Czerny-Turner spectrometer, achieving high wavelength resolution and accuracy in a compact size. In addition, the use of a high-precision, high-sensitivity line sensor makes it ideal for high-speed measurement.
Moreover, the spectrometer features a light guide fiber input method, making it extremely easy to handle.

OCM-510 appearance

Features

  • Compact spectroradiometer integrating a small high-performance spectrometer and detector
    OCM-510M : MOS line sensor enables high dynamic range spectroradiometry OCM-510C : Cooled back-illuminated FFT-CCD enables extremely low-light measurement (approx. 100x sensitivity of MOS)
  • Minimum exposure time of 5 ms (MOS type)
    (Higher speed and high-brightness light input capability compared to OCM-310)
  • Selectable amplifier gain (1/2/4x) after sensor readout
  • Easy measurement with light guide fiber
  • External sync signal output for pulsed light measurement
  • Models with mechanical shutter available (standard on OCM-510M or C-SU)
  • USB or LAN interface for PC communication
  • Spectrometer control and color calculation output function※1 (via PC software)
※1 Separate software for spectrometer control and color calculation is available.
OCM-510 can also perform color calculation internally.
For details regarding software, please contact us separately.

Spectrometer Characteristics

The OCM-510M/C is a compact, high-wavelength-accuracy polychromator employing a Czerny-Turner optical configuration. The Czerny-Turner spectrometer consists of an entrance slit, two concave mirrors, and a plane grating. It directs parallel light onto the plane grating, and the diffracted parallel light is focused by a concave mirror onto a photodiode array detector. This symmetrical optical arrangement with respect to the grating corrects curved spectral lines back to vertical, thereby improving wavelength resolution.

Furthermore, the plane grating features a blazed holographic grating* with extremely low periodic errors and ghosting.

Blazed holographic grating
*Blazed holographic grating
A grating fabricated by holographic exposure using two-beam interference of a laser, then blazed by ion beam etching. Although it is a holographic grating, its diffraction efficiency is not inferior to ruled gratings, and it exhibits low stray light.

Light Trigger Function

By using the light trigger signal (synchronous output for pulsed emission) from the OCM-510 to light an external source, high-speed and accurate pulsed measurement is possible.

Light trigger function image

Specifications by OCM Model

OCM series specification table
  1. Linearity at 450 nm, 500 nm, 650 nm for 2:1 incident light
  2. Linearity at 450 nm, 500 nm, 650 nm for 10:1 incident light
  3. Twice the standard deviation of 10 repetitions under condition 1
  4. Twice the standard deviation of 10 repetitions under condition 2
  5. Ratio with/without sharp-cut glass filters at 450 nm, 500 nm, 600 nm using a tungsten bulb as light source
* These items are supplementary information corresponding to the specification table above.